Electric Handpiece | BJ6200 |
Electric: | Yes |
Multifunctional: | Yes |
Weight (with battery): | ≤1000 Grams |
Speed: | 0-15000 rpm |
Power source: | Mains and Battery |
Sterilizable: | Autoclavable |
Battery: | BJ6001 |
Technology: | Ni-Mh |
Insert/detachment method: | The telescopic |
The blocking mechanism: | Yes |
Autoclavable Aseptic Transfer Kit: | Yes |
Run time with battery: | 30-40 min |
Simultaneously charged batteries: | 4 |
Battery charge status: | Indicator light |
Electric Power Supply: | BJ6006+ BJ6007+BJ6008+BJ6009 |
Handpiece main power cord: | 2 pcs |
Sterilizable method: | Autoclavable |
Handpiece main power cord length: | 3500mm - 5000mm |
Sagittal saw attachment: | BJ6201 |
Oscillations (osc/min) | 0-15000 |
Operation(V) voltage | 9.6 |
Sterilization | 155°C |
Noise(db) | ≤75 |
Acetabulum reaming drill attachment: | BJ6207B |
Speed(r.p.m) | 0-300 |
Torque(Nm) | 7.8 |
Operation(V) voltage | 9.6 |
Sterilization | 155°C |
Noise(db) | ≤40 |
Canulate drill attachment | BJ6203B |
Speed(r.p.m) | 0-630 |
Torque(Nm) | 3.6 |
Operation(V) voltage | 9.6 |
Sterilization | 155°C |
Noise(db) | ≤40 |
1.Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the |
temperature is over 135 ° C ,the motor will be damaged. |
2.Normal bone drill can only hold drill bits less than 6mm,acetabulum |
reaming drill with big torque,rpm is around 300,canulate drill is used |
with nail or K wire ,rpm is around 600. |
3.The battery need to be get out of the pwer tool after surgery. Otherwise |
the small automatic discharge of the electric will make the life of the |
battery short. |
4.The battery cannot be sterilized ,it is not autoclavable. |
5.The cranial bur must be washed right after the surgery by purger in |
order to aviod the blood stick in the bur,then we should make it dry. |
6.The battery space in the handpiece of the tool should not have any |
water,after sterilization. |
7.The cranial bur blade is made of medical stainless steel,in order to |
avoid broken,so it is flexible and not very hard,it is easy to be attrite, |
we suggest each blade only use for 1 surgery. |