Customization: | Available |
---|---|
Type: | Drill and Saw |
Application: | Orthopedic, Neurosurgery |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Power tool |
Sagittal saw (BJ2101) |
Oscillations (osc/min) |
0-15000 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤75 |
Power tool |
Bone drill(BJ2102) |
Speed(r.p.m) |
0-1000 |
Torque(Nm) |
2.9 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤40 |
Power tool |
Canulate drill(BJ2103) |
Speed(r.p.m) |
0-630 |
Torque(Nm) |
3.3 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤40 |
Power tool |
Cranial drill(BJ2104) |
Speed(r.p.m) |
0-1000 |
Torque(Nm) |
2.9 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤40 |
Power tool |
Cranial bur(BJ2105) |
Speed(r.p.m) |
0-40000 |
Sterilization |
135°C |
Noise(db) |
≤40 |
Power tool |
Sternum saw(BJ2106) |
Oscillations (osc/min) |
0-15000 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤70 |
Power tool |
Acetabulum reaming drill(BJ2107) |
Speed(r.p.m) |
0-280 |
Torque(Nm) |
7.8 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤40 |
Power tool |
Reciprocating saw(BJ2109) |
Oscillations (osc/min) |
0-15000 |
Operation(V) voltage |
14.4 |
Sterilization |
135°C |
Noise(db) |
≤70 |
1.Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the |
temperature is over 135 ° C ,the motor will be damaged. |
2.Normal bone drill can only hold drill bits less than 6mm,acetabulum |
reaming drill with big torque,rpm is around 300,canulate drill is used |
with nail or K wire ,rpm is around 600. |
3.The battery need to be get out of the pwer tool after surgery. Otherwise |
the small automatic discharge of the electric will make the life of the |
battery short. |
4.The battery cannot be sterilized ,it is not autoclavable. |
5.The cranial bur must be washed right after the surgery by purger in |
order to aviod the blood stick in the bur,then we should make it dry. |
6.The battery space in the handpiece of the tool should not have any |
water,after sterilization. |
7.The cranial bur blade is made of medical stainless steel,in order to |
avoid broken,so it is flexible and not very hard,it is easy to be attrite, |
we suggest each blade only use for 1 surgery. |